Temperature and humidity testing on the bench: 50 or 80 litres, from -86 to +150 °C and 20 to 98% RH, at 65 dBA or less.
| Model | T-50 | T-80 |
|---|---|---|
| Internal Dimension (mm) | 320 × 350 × 450 | 400 × 400 × 500 |
| Overall Dimension (mm) | 800 × 1050 × 950 | 900 × 1100 × 1000 |
| Capacity | 50 L | 80 L |
| Temperature Range | A: -20 °C to +150 °C | |
| Low Type | B: -40 °C; C: -70 °C; D: -86 °C | |
| Temperature Fluctuations | ±0.5 °C | |
| Temperature Deviation | ±2.0 °C | |
| Heating Rate | 3 °C / min | |
| Cooling Rate | 1 °C / min | |
| Controller | Programmable color LCD touch screen controller, Ethernet, USB | |
| Cooling System | Mechanical compression refrigeration system | |
| Safety Device | over-temperature protection; over-current protection; Refrigerant high-pressure protection; Earth leakage protection | |
| Exterior Material | Steel Plate with protective coating | |
| Interior Material | SUS304 stainless steel | |
| Standard Configuration | 1 Cable hole (Φ 50) with plug; 2 shelves | |
| Power Supply | 220 V 50 Hz | |
| Model | TH-50 | TH-80 |
|---|---|---|
| Internal Dimension (mm) | 320 × 350 × 450 | 400 × 400 × 500 |
| Overall Dimension (mm) | 800 × 1050 × 950 | 900 × 1100 × 1000 |
| Capacity | 50 L | 80 L |
| Temperature Range | A: -20 °C to +150 °C | |
| Low Type | B: -40 °C; C: -70 °C; D: -86 °C | |
| Humidity Range | 20% to 98% RH | |
| Temperature Fluctuations | ±0.5 °C | |
| Temperature Deviation | ±2.0 °C | |
| Heating Rate | 3 °C / min | |
| Cooling Rate | 1 °C / min | |
| Humidity Deviation | ±2.5% RH | |
| Controller | Programmable color LCD touch screen controller, Ethernet, USB | |
| Cooling System | Mechanical compression refrigeration system | |
| Safety Device | Humidifier dry-combustion protection; over-temperature protection; over-current protection; Refrigerant high-pressure protection; Water shortage protection, Earth leakage protection | |
| Exterior Material | Steel Plate with protective coating | |
| Interior Material | SUS304 stainless steel | |
| Power Supply | 220 V 50 Hz | |
Tell us the specimen, the test standard and the temperature range, and we propose the chamber. Request a quote
INTEON Benchtop Climate Test Chamber: Temperature and humidity testing on the bench: 50 or 80 litres, from -86 to +150 °C and 20 to 98% RH, at 65 dBA or less.